Starting the Polishing/Etching
Process
Stopping the Polishing/Etching
Process
Movipol-3
Instruction Manual
Make sure that the correct type and amount of electrolyte has
been filled in the electrolyte container.
Connect the anode to the sample by means of the clamp or
magnet supplied with the apparatus.
Set the polishing voltage and polishing time.
Set the etching voltage and etching time.
Activate the pump motor by switching O/Charge - l/Pump to
bottom position.
Take the pencil from the pencil holder and place it on the spot
which is to be prepared.
Press the pencil against the surface ensuring that the polishing
chamber is in total contact with the surface.
Keep the pencil perpendicular (at a angle of 90°) to the surface.
The electrolyte will now begin to circulate.
Take care that no air is drawn around the periphery of the polishing
Press the button START
there is electric connection from the cathode in the pencil
through the electrolyte to the anode.
If the electric circuit is closed the timer starts the process.
Only press the start button when the noise from the pump indicates that the
flow through the pencil is constant.
Read the ammeter to check if the correct amperage has been
obtained.
When both the Polishing and Etching Indicators are switched off
the process has been completed and the pencil should be
placed in the pencil holder.
When the pencil has been placed in the holder, switch off the
pump motor by switching the O/charge - l/Pump to O/Charge.
Clean the polished spot with water/alcohol and dry carefully.
The polishing/etching process can be stopped at any time during
operation by setting the adjusting knob for polishing/etching time
to zero.
IMPORTANT
chamber.
. A "sensor circuit" records whether
IMPORTANT
9