Name
SOCKET
ANODE
Safety Functions
Cut-out of Pump Motor
Current Limiter Function of the
Polish/Etch Circuit
Cut-out of Polish/Etch Circuit
Cut-out in case of under-voltage
Mounting the Container
Dismounting the Container
Movipol-3
Instruction Manual
Function
For connecting the anode wire to
the control unit.
In case of higher load than 1 A the cut-out function is activated. The
Voltage Indicator for Pump Motor will light red. The apparatus can be
restarted when the error has been eliminated.
In case of a voltage range of 70-90 V the current is limited
corresponding to the maximum effect of the transistors (140 VA).
In case of a voltage range of 2-70 V at 2 A the cut-out function is
activated and an alarming signal is heard. When the error has been
corrected the apparatus can be restarted. In case of overload the
circuit is cut out and the Voltage indicator for Pump Motor lights red.
The apparatus can be restarted after cooling when the indicator
lights yellow.
Under-voltage may occur using insufficiently charged batteries. The
cut out function for under-voltage will disconnect the process and an
alarm signal is heard.
Place the pump in the container.
Press down the black knobs and turn them until the markings
turn towards the pump.
Check that the container is firmly locked.
Press the two black knobs down and turn them so that the
markings turn away from the pump, a click is heard and the
pump is released from the container.
Name
SOCKET
FOR
EXTERNAL
ETCHING
Function
For connecting the cathode
when external etching
operations should be carried
out.
7