Using the Controls

Front Panel Controls

Name
O/CHARGE
I/PUMP SWITCH
START BUTTON
ADJUSTING KNOB FOR
POLISHING VOLTAGE
ADJUSTING KNOB FOR
ETCHING VOLTAGE
POLISHING TIME
ETCHING TIME
6
Movipol-3
Instruction Manual

2. Operation

Function
When switched to O/Charge the
batteries are being charged if
the apparatus is connected to
mains. When switched to l/Pump
the pump motor starts operating.
Push button for starting the
polishing/etching process.
Adjusting knob for setting the
voltage for polishing. It can be
set in the range of 10-90 V. The
voltage range of 60-90 volt DC is
marked with a voltage danger
warning.
Adjusting knob for setting the
voltage for etching.
Adjusting knob for setting the
time of polishing. When set to ∞
the process continues until the
knob is set to zero.
Adjusting knob for setting the
time of etching. When set to ∞
the process continues until the
knob is set to zero.
Name
AMMETER
CHARGE INDICATOR
VOLTAGE INDICATOR
FOR PUMP MOTOR
POLISHING
INDICATOR
ETCHING INDICATOR
SOCKET PUMP/
CATHODE
Function
For reading the amperage
during the polishing/etching
process. The range of 0-2 is
used for polishing. The range of
0-0.2 A is used for etching.
Changing between the ranges is
done automatically when the
timers switch from polishing to
etching.
Indicates the battery charging
level. When connected to 42V
AC possible readings are as
follows:
Battery Charging
Level
O/
Charge
Medium charged Yellow
Fully charged
Green
Not Charged
Red
Indicates whether the pump is
supplied with sufficient voltage:
0: The apparatus is switched off.
Yellow: Pump motor in operation
at sufficient voltage.
Red: Pump motor blocked due
to overload of pump motor or too
low voltage. Electrical circuit
switched off due to overload.
The Polishing Indicator lights red
when the polishing process is
running and is switched off when
the process is finished.
The Etching Indicator lights red
when the etching process is
running and is switched off when
the process is finished.
For connecting the pump motor
and cathode to the control unit.
I/
Pump
Green
Green
Red