Error message
TegraPol
NVRAM initialised
Frequency inverter communication
error
Frequency inverter heat sink temp.
too high
Frequency inverter internal temp.
too high
Frequency inverter unknown error:
1
Frequency inverter type mismatch
error
Motor temperature too high
Motor temp. sensor disconnected
Cannot read keyboard
TegraPol with TegraForce
TegraForce must be down for
preparation
TegraForce not locked
TegraForce must be up for manual
preparation
TegraForce must be up when
cleaning
Check water expeller on
TegraForce
TegraPol-11/ -15
TegraForce-1
Instruction Manual
7. Trouble-shooting
Explanation
Default settings have been set for
all non-volatile parameters.
Internal frequency inverter error.
Fabrication error, incorrect motor
controller mounted in device.
Action required
None, message will be removed
after a few seconds
Turn device off and on, and try
again. If problem persists call a
Struers service technician.
Turn device off and let it cool down
for a few minutes. Turn device on
and try again. If problem persists
call a Struers service technician.
Turn device off and let it cool down
for a few minutes. Turn device on
and try again. If problem persists
call a Struers service technician.
Turn device off and on, and try
again. If problem persists call a
Struers service technician.
Call a Struers service technician.
Stop process and let device cool
down for a few minutes. Try again. If
problem persists device call a
Struers service technician.
Call a Struers service technician.
Call a Struers service technician.
Press up/down button before
starting process.
Lock TegraForce with handle before
starting process.
Press up/down button before
starting manual process.
Press up/down button before
starting cleaning.
Press the release outlet valve to
drain the water/oil filter.
37