Evidently, the default settings can be bypassed and the "Controlled
Venting Procedure" can be eliminated, as explained in a previous
section of this manual.
Additionally, the vent port on the envelope can be used to connect an
external vent valve: unscrew the threaded plug and screw the vent
valve into the pump to vent pump and system.
The last described venting modalities should be avoided because they
may cause uncontrolled venting. Please contact Agilent if you cannot
operate in the instructed manner.
Uncontrolled venting may seriously damage the pump and void the pump
CAUTION!
warranty. Use the Agilent "Controlled Vent Procedure" or contact Agilent.
Figure 47
Turbo-V 3K-G Pumping System User Manual / 87-901-011-01
1
Purge gas line
2
Pressure regulator
3
Purge and vent port
4
Forevacuum pump
5
Turbopump
6
Water line
Purge and vent layout
Technical Information
Pump Purging and Venting
VACUUM TECHNOLOGIES
(C)
17
365/374