In remote mode you can open the purge valve (N.C.) connecting
pin10 of J1 to pin 9, or supply-ing 24 Vdc between pin 10 and pin 15.
The same thing occurs for the vent valve with pin 12.
Note that the vent and purge valve can also be controlled by means of
the serial connection (see the following diagram for details).
1=Remote-I/O ( def )
Pin-10 Low -> Purge Close
Pin-10 High -> Purge Open
Pin-12 Low -> Vent Close
Pin-12 High -> Vent Open
win.122=0 Vent Close (def)
win.122=1 Vent Open
win.145=0 Purge Close
win.145=1 Purge Open (def)
Fail -> Vent and Purge =
Closet
Power Fail -> Purge and
Vent = Closed
Figure 17
Turbo-V 3K-G Pumping System User Manual / 87-901-011-01
1 "OnCommand"
Wi
146
0=Serial
Command
Start
Ramp
Status
Speed
100 Hz
Open
Purge
Close
Vent
Technical Information
0="Controlled Vent Procedure"(def)
Wi
125
Win.126 set the "Vent Valve Opening Delay",
from 0 to 65535, 1bit=0.2s, def=0.2s=1bit
Win.147 set the "Vent Valve Open Time",
from 0 to 65535, 1bit=0.2s, def=0=infinite
Stop
Normal
Close
(C)
Interconnections
Braking
Stop
Open
Controlled
Win 126
Win 147
Venting
317/374