Agilent Technologies Turbo-V 3K-G 969-8888 Notice De Mode D'emploi page 354

Table des Matières

Publicité

17
Technical Information
Profibus Message Mapping
Offset
Size
Unit
Byte
Byte
Res
1
2
Hz
3
2
Hz
Tab. 14
Input buffer (15 words)
Offset
Size
Unit
Byte
Byte
Res
0
2
mA
2
2
W
4
2
W
6
2
Hz
8
2
Hz
10
2
-
12
2
-
14
2
-
16
2
min
18
2
hours
20
1
V
21
1
°C
22
1
°C
354/374
Range
Description
500 -
High Speed Value
530
Set the high rotational speed, default = 530
500 -
Low Speed Value
530
Set the low rotational speed, default = 500
Range
Description
0-10000
Current. Motor current consumption
0-750
Power. Motor power adsorption
0-750
Power Limit Applied. Maximum allowable power
0-530
Driving Frequency. Read the rotational speed of the pump
0-530
Target Frequency. Read the target rotational speed
-
Flow Meter Alarm Time
Specific option, not available on standard models
-
Flow Meter Alarm Event
Specific option, not available on standard models
0-65536
Cycle Number. Number of cycles (start and stop) done by the
pump
0-65536
Last Cycle Time. How long the last cycle lasted
0-65536
Pump Life. Total pump running time
0-100
Voltage
Voltage provided to the motor
0-70
Pump Bearing Temperature
Read the pump upper bearing temperature.
If > 60 °C the controller goes in fail and the pump is stopped
0-45
Pump Body Temperature
Read the pump body temperature close to the water circuit.
If > 42 °C the controller goes in fail and the pump is stopped
Turbo-V 3K-G Pumping System User Manual / 87-901-011-01
Win
Serial
120
117
Win
Serial
200
201
155
203
120
213
215
301
300
302
201
204
216