Do not install or use the pump in an environment exposed to
atmospheric agents (rain, snow, ice), dust, aggressive gases, or in
explosive environments or those with a high fire risk. During
operation, the following environmental conditions must be respected:
In the presence of magnetic fields the pump must be protected using
a ferromagnetic shield. See the appendix "Technical Information" for
detailed information.
The Turbo-V 3K-G must be connected to a primary pump (see
"Technical Information").
The Turbo-V 3K-G can be installed in any position. Fix the Turbo-V
3K-G in a stable position connecting the inlet flange of the
turbopump to a fixed counter-flange capable of withstanding a torque
of 33000 Nm around its axis.
The turbopump with ISO F inlet flange must be fixed to the vacuum
chamber by means of bolts with a strength class of 12,9
(σr = 1200 N/mm
of steel bolts and the relevant fixing torque as recommended by
Agilent.
Tab. 1
FLANGE
ISO 250 F
WARNING!
The Turbo-V 3K-G cannot be fixed by means of its base. The system can only
be fixed through its ISO 250F flange.
Turbo-V 3K-G Pumping System User Manual / 87-901-011-01
maximum pressure: 2 bar above atmospheric pressure
air temperature: from +5 °C to +35 °C
relative humidity: 0 – 95 % (non-condensing)
2
). The following table shows the necessary number
FIXING DEVICE
M10 bolts
Instructions for Use
Installation
N.
FIXING TORQUE
12
30 Nm
(C)
16
289/374