Tab. 13
Output buffer (5 bytes)
Offset
Size
Unit
Byte
Byte
Res
0
1
Turbo-V 3K-G Pumping System User Manual / 87-901-011-01
Range
Description
B7 – Start/Stop
0 = Stop
1 = Start
B6 – Low Speed
0 = off (set the "target speed"(In Buff byte-8) at" high
speed" (Out. Buff. byte-1))
1 = on (set the "target speed" (In Buff byte-8) at "low
speed" (Out. Buff. byte-3))
B5 – Soft Start
0 = off (this function is active at first pump run-up and can't
be deactivated by the user, after first run-up activate this
function only if the pump has been stopped for at least 3
months)
1= on
B4 – Active Stop
0 = off (after a stop command the controller stop driving the
motor, the pump will decrease its speed by friction with gas
load)
1 = on (after a stop command the controller will brake the
pump using the motor, activate this function only if you don't
use the vent valve and for few cycles at week)
B3 - Vent Valve On/Off
0 = close, 1 = open
(Valid only if in Parameter buffer Byte-15-B1 = 1, and Byte-
15-B2 = 0)
B2 – Purge Valve On/Off
0 = close, 1 = open, (Valid only if in Parameter buffer Byte-
15-B1 = 1, and Byte-15-B2 = 0)
B1 – Heather Jacket Enable.
Specific option, not available on standard models
B0 – Stop Speed Reading
0 = no (after a stop command the user can't know the real
pump speed)
1 = yes (after a stop command the user can know the real
rotor speed; see in Buff Byte Offset 6)
Technical Information
Profibus Message Mapping
(C)
17
Win
Serial
000
001
100
107
122
145
112
167
353/374