Operation
At predefined intervals, the MoniTorr will also acquire the data
related to the vibration spectrums, power and temperature in
addition to other data detected by the Turbo controller. All the data
must be transferred to a PC (by means of the "File Download"
program) where are stored and where can be analyzed by Agilent's
specialised personnel.
Maintenance
The MoniTorr card doesn't require any maintenance. Any
intervention must be carried out by authorised personnel.
In case of failure, the Agilent Repair Service is available or Advanced
Exchange Service that provides you with a regenerated controller in
replacement of the faulty one.
MoniTorr Installation on the Network and on the
Main PC
The RS-485 interface allows connecting more than one system in
network (up to 32). In this way it is possible to download all the
spectrums data with a single PC through a protocol converter (P/N
9699258 and 9699259) which operate as a gateway from 485 to 232
link.
The controller provides the serial address of the MoniTorr option.
Every controller must therefore have a different serial address so as
to avoid conflicts on the network. Every controller has "0" default
serial address. So the user must set the controller's address.
The following figure shows the connection of the MoniTorr to the
RS485 network.
Turbo-V 3K-G Pumping System User Manual / 87-901-011-01
Technical Information
MoniTorr
(C)
17
333/374