Agilent Technologies Turbo-V 3K-G 969-8888 Notice De Mode D'emploi page 293

Table des Matières

Publicité

WARNING!
When power is supplied and factory default 15 pin mating connector is
inserted, the Turbo-V 3K-G will start automatically.
The controller is furnished already mechanically and electrically connected to
CAUTION!
the pump. Detaching of the controller from the pump must be carried out by
authorized Agilent Vacuum Technologies personnel only.
When you run the Turbo-V 3K-G for the very first time, the control unit
NOTE
automatically starts the system with a special procedure which protects the
bearings from possible damages (SOFT START). The system is launched step by
step at full speed in a time variable from 10 mins up to 1 hour. After the system
has reached the full speed, the soft start procedure is disabled and following
starts are performed in the normal way.
In order to maintain the IP-54 protection level you must use only the connectors
NOTE
provided with the pump. Use connectors PN: 969-9957 or 969-9958 for the power
cable and fix it to the controller by means of the suitable retaining bracket (see
the following figure). Use this cable and plug together with a properly grounded
plug to prevent electric shocks and to comply with EC Standards.
Turbo-V 3K-G Pumping System User Manual / 87-901-011-01
Instructions for Use
(C)
16
Use
293/374

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

Turbo-v 3k-g 969-8889

Table des Matières