17
Technical Information
Profibus Option
Installation
This option is factory installed.
The profibus connector pin-out is reported in the following figure.
Please in order to obtain best performance use profibus certified
cables.
To maintain the IP-54 protection level, use the appropriate connector provided
CAUTION!
with the pump.
Figure 40
Configuration
Before power-on the controller set the Profibus address by the 2
rotary switches.
Up to 126 different addresses (from 0 to 125) can be selected. The
address value sets by the 2 switches is expressed in hexadecimal
notation, so value from 00 to 7D are permitted; this setting is read by
the interface during the power-on phase, so any change of the switch
position after the power-on is ignored.
346/374
ProfiBus connector
Turbo-V 3K-G Pumping System User Manual / 87-901-011-01
(C)