Figure 15
FAULT: this relay output signal is closed when any system fault
condition is detected. To detect the type of failure the serial protocol
is needed (see the para. "RS232-485 Comunication Description").
SET POINT: this relay output signal is enabled when the reference
quantity chosen (frequency, current or time) is higher than the set
threshold. The signal can be "active close", or "active open". Moreover,
if the reference quantity is the frequency or the current drawn, it is
possible to set the hysteresis (in % of the threshold value) to avoid
bouncing.
For example:
Turbo-V 3K-G Pumping System User Manual / 87-901-011-01
reference quantity: frequency (window 101=0)
threshold: 500 Hz (window 102=500)
hysteresis: 1% (window 103=1)
activation type: "active close" (window 104=1)
Technical Information
Interconnections
(C)
17
315/374