7.2

Measuring the diluted oxygen

7.2.1

Before the measurement

7.2.2

Measuring

Measuring with the O2 Probe (option)
Connect the O
Switch the meter on.
The O
reading is displayed.
2
NOTE!
If the O
probe is connected to the device for
2
the first time or after a longer period of time, it
may take 10 minutes for the value to stabilize.
With regular use and the "fast" setting, the value
will stabilize in a few seconds. The settings for
the stabilization time can be made in the set-
tings menu, see chapter 10.6.
Run a 100 % calibration on air, see chapter
11.
Dip the probe into the sample. Make sure
the electrode is totally immersed when you
stir it.
Wait until the reading is stabilized.
Press the O
-key to switch between the
2
units % and mg/l.
-Probe to the meter.
2
EN
63