7.2
Measuring the diluted oxygen
7.2.1
Before the measurement
7.2.2
Measuring
Measuring with the O2 Probe (option)
Connect the O
•
Switch the meter on.
•
•
The O
reading is displayed.
2
NOTE!
If the O
probe is connected to the device for
2
the first time or after a longer period of time, it
may take 10 minutes for the value to stabilize.
With regular use and the "fast" setting, the value
will stabilize in a few seconds. The settings for
the stabilization time can be made in the set-
tings menu, see chapter 10.6.
•
Run a 100 % calibration on air, see chapter
11.
•
Dip the probe into the sample. Make sure
the electrode is totally immersed when you
stir it.
Wait until the reading is stabilized.
•
Press the O
-key to switch between the
•
2
units % and mg/l.
-Probe to the meter.
2
EN