Position of the electrode
Fig. 46: Electrode
Cleaning the electrode
Changeing the membrane set
Measuring with the O2 Probe (option)
Check the central of the electrode surface. It
should be clean in the working area (see
photo).
Clean the electrode if too many residues
have been found on electrode working area.
(Too many residues attached might disturb
the chemical reaction during measuring.)
Rotate to unscrew the membrane set from
the probe.
CAUTION!
Any electrolyte still present may leak out. Hold
the electrode down.
Clean the probe with a soft fiber cloth or if
necessary with the finger nail.
Use a knife blade to scrape lightly on the
electrode to remove the white residue as
below image. Do not scrape the "surface of
working area" by knife blade to prevent
damage.
Use clean water to wash away the remain-
ing impurities.
NOTE!
Under normal use, the 2 Replacement Mem-
brane Sets enclosed in the package of the O
Probe will be enough to cover the replacement
need for the entire O
Replace membrane set when any of the
Probe life time.
2
EN
-
2
61