Temperature measurement connection:
For external temperature measurement, any type K
thermocouple can be connected , e.g. a tempera-
ture clamp or a temperature probe.
Insert the temperature plug into the tempera-
•
ture socket on the underside of the unit.
•
Immediately after switching on, as soon as the
battery symbol has appeared in the display
and before the measured value display ap-
pears, double-click with the button.
This will take you to the setting mode.
The display shows „Press unit".
To select the pressure unit, press the button
•
briefly.
The pressure units mbar, bar, in
mH
O, Pa, hPa, kPa and mPa appear one after
2
the other.
•
Confirm the selected unit with a double click.
The display shows „reso" (resolution).
You can now set the displayed resolution for the
pressure. The resolution refers to the previously
selected pressure unit. There are 3 different reso-
lutions to choose from.
Press the button briefly to select the desired
•
resolution.
Confirm the selection with a double click.
•