11.2.3 Prior to calibration

Fig. 58: EC calibration
tion values from other measurement ranges that
have not been recalibrated are retained
Clean the electrodes using a soft brush.
NOTE!
The probe guard must not be removed during
the calibration procedure.
Place the probe tip in demineralized or
distilled water for approx. 30 min.
Empty the calibration solution into a clean
container.
Power on the measuring instrument and
immerse the probe in the calibration solu-
tion; ensure the electrodes are completely
covered.
Stir lightly for approx. 1 minute. Lightly tap
the probe to remove any bubbles adhering
to the electrodes.
CAUTION!
Make sure the electrodes are completely cov-
ered by the calibration solution, but that they do
not contact the bottom of the container (screw
on the probe guard!).
To enter the calibration mode press and
hold the CAL key for approx. 2 seconds.
As soon as the device has measured the
conductivity of the calibration solution, the
reading flashes in the display
If the displayed value deviates from the
setpoint value, use the arrow keys to set the
value.
NOTE!
If the arrow key is held down, the value can be
set more quickly
Calibration
EN
73