Fault
Possible cause
message
E02
The measured value is below the
measurement range.
E03
The measured value is above the
measurement range.
E04
Data error
E13
Error during pH calibration.
E16
Cell constant of the EC probe is outside
the measurement range
Remedy
Place the probe you are
using into normal water at
room temperature for 30
minutes until the message
E02 is no longer displayed.
If that fails to resolve the
problem, recalibrate the
probe.
Place the probe you are
using into normal water at
room temperature for 30
minutes until the message
E03 is no longer displayed.
If that fails to resolve the
problem, recalibrate the
probe.
Check if the measured tem-
perature is above the meas-
urement range.
the probe you are using in-
to normal water at room
temperature for 30 minutes
until the message E04 is no
longer displayed.
Run another pH calibration
with fresh calibration solu-
tion.
Power the measuring in-
strument on and off several
times in succession.
Recalibrate the EC probe.