1.5.1
Storing the pH probe
1.5.2
Storing the EC probe
1.6
Disposal
1.7
Manufacturer's address
CAUTION!
Ensure the glass bulb of the probe is kept moist at
all times. During storage, the glass bulb must be
covered by KCI electrolyte storage solution.
A graduated beaker to fill a measured quantity of
KCI storage solution is included in the scope of
supply.
Store the probe with the transparent cap, but
•
without the probe guard.
Fill the transparent cap with KCI electrolyte
•
storage solution before storing.
Alternatively, it is possible to use tap water as stor-
age solution.
CAUTION!
Never use distilled water as a storage solution!
CAUTION!
Make sure the electrodes do not come into con-
tact with the KCI electrolyte solution (storage solu-
tion for the pH probe).
Store the probe in a dry location with the
•
probe guard screwed on.
Do not dispose of electronic equipment along with
household waste. Dispose of electronic equipment
in accordance with applicable environmental regu-
lations. Damaged rechargeable batteries are con-
sidered hazardous waste. Dispose of damaged or
used rechargeable batteries at the collection
points provided.
Wöhler Technik GmbH
Wöhler-Platz 1, 33181 Bad Wünnenberg,
Germany