Faults

12
Faults
Fault
It is not possible to power
on the device.
pH reading not stable
76
the meter
- after changing the membrane set
- after changing the battery
Possible cause
Pressed the power
on button too briefly.
Batteries are flat or
battery connection
fault
Low battery power
Test pH probe too
old
Remedy
Power on again.
Place the batteries correctly
in the compartment or re-
place
Replace batteries
Never use old and new
batteries at the same time
Replace pH probe.