O 2 -Calibratation; Calibration Interval; Calibration - Wohler WA 335 Instructions De Service

Appareil d'analyse de l'eau
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

11.3 O
-Calibratation
2

11.3.1 Calibration interval

11.3.2 Calibration

Fig. 60: O2- calibration
Error Message
Stabilization period
Perform an O
2
measurement and after replacing the
membrane set.
NOTE!
There is no need to remove the anti-collision
hood before the calibration.
Connect the O
switch it on.
Press O
-key to select %.
2
Hold the probe in air and wait a few minutes
until the reading is stabilized.
Press and hold CAL-key for 2 seconds for
100 % saturation calibration.
CAL Icon is flashing in the LCD.
When the reading is stabilized after some
seconds, press ENTER.
or
You can stop the calibration by pressing
CAL again.
After the calibration the displayed value should
be between
Nach der Kalibrierung sollte der angezeigte Wert
zwischen 99,0 % und 101,0 % liegen.
If the meter cannot perform a correct calibration,
ERR will flash on the display.
Possible reasons:
- electrolyte level too low
- defective probe
- insufficient adjustment time
Under the following conditions, the meter
requires an adaptation time of approx. 1 hour.
Switch the device on and leave it for about
one hour,
- when the O
probe has just been connected to
2
–calibration before each
- Probe to the meter and
2
Calibration
EN
75