17
Technical Information
TV 1001 SEM Controller Installation
Bottom Mounting
The L-shaped bracket (P/N 9699298) is available as an option.
NOTE
See the following figure:
1
2
3
4
5
6
7
Figure 43
282/296
Turn the pump upside-down;
Place the three fixing blocks with the through holes toward the inside, and
fix them with the three M6 screws to the pump bottom;
Place the controller on the blocks, with the pump cable toward the pump
body;
Fix the controller to the blocks by means of the three socket head screws
M5;
Turn the pump again;
Plug the line card and connect the pump cable to the pump;
Connect the mating connector with the jumper on the interlock signal to
start the pump.
Turbo