Connection B – Fore-Vacuum Pump
A flange KF 40 NW is available to connect the TV 1001 SEM pump to the fore-
vacuum pump. A hose or vacuum approved pipe can be used. If a rigid pipe is
used, any vibration generated by the mechanical pump must be eliminated
through the use of bellows.
The TV 1001 SEM pump is characterized by its high compression ratio also for
NOTE
oil vapors. When using a mechanical oil-sealed pump, it is advisable to install a
suitable trap between the turbopump and the fore-vacuum pump in order to
prevent oil backstreaming.
Connection Configurations
Figure 46
TV 1001 SEM User Manual / 87-900-973-01 (H)
Technical Information
TV 1001 SEM Controller Installation
17
285/296