Inlet Screen Installation
Figure 28
The inlet screen mod. 9699316 prevents the blades of the pump from being
damaged by debris greater than 0.7 mm diameter.
The inlet screen, however, does reduce the pumping speed by about 10%.
The inlet screen is fitted in the upper part of the pump, as shown in the figure.
Figure 29
The screen can be mounted on each pump.
The screen can be removed as shown in the following figure.
TV 1001 SEM User Manual / 87-900-973-01 (H)
Technical Information
Inlet Screen Installation
17
271/296