Instructions for Use
TwisTorr 305-IC Mounting
TwisTorr 305-IC Mounting
WARNING!
Failure to comply with these installation instructions could result in the pump
detaching from the system in the event of a rotor failure, which could cause
property damage or serious injury or death.
The TwisTorr 305-IC can be installed in any position. Fix the TwisTorr 305-IC in
a stable position, mounting the inlet flange of the turbopump to the system
counter-flange, with a connection capable of withstanding a torque of 1000 Nm
around its axis.
The turbopump with ConFlat and ISO K inlet flange must be fixed to the vacuum
chamber by means of the appropriate Agilent hardware.
For further details see the "Installation of ConFlat and ISO K" section.
The TwisTorr 305-IC cannot be fixed by means of its base
CAUTION!