Specifications subject to change without notice.
Harvester
Dimensions: .......................................................................................................................5.7" W x 3.7" H x 6" D (14.5 x 9.4 x 15.2 cm)
Weight: ....................................................................................................................................................................................... ~0.2 lbs (~0.09 kg)
Vacuum Head
Dimensions: ......................................................................................................................5.5" W x 3.9" H x 1.3" D (13.9 x 9.9 x 3.3 cm)
Weight: ....................................................................................................................................................................................... ~0.6 lbs (~0.27 kg)
Control Unit
Dimensions: ............................................................................................................ .10.6" W x 9.3" H x 9.4" D (26.9 x 23.6 x 23.9 cm)
Weight: ....................................................................................................................................................................................... ~7.3 lbs (~3.31 kg)
Control Unit Basket Accessory
Maximum weight rating........................................2 lbs (the approximate weight of 4 individual harvester packages)
Electrical
Internal Power Supply Input: ...............................................................................................................100-240 VAC 3.2A 50 / 60 Hz
Internal Power Supply Output: ............................................................................................................................................ 24 VDC, 10.4 A
Power Inlet Fuse
Type: 5x20mm Time Lag glass cartridge
Current Rating: 5A
Voltage Rating: 250V
Operative Speed(s): 600ms ≤ t ≤ 10s @ 13.75A
Breaking Capacity: 150A @ 250V
Storage / Transport Conditions
Temperature Range: ......................................................................................................................................... 0°F (-18°C) to 140°F (60°C)
Relative Humidity Range: ..............................................................................................................................15 - 95% non-condensing
Operating Conditions
Temperature Range: ...........................................................................................................................................58°F (14°C) to 95°F (35°C)
Relative Humidity Range: ..............................................................................................................................20 - 80% non-condensing
Atmospheric Pressure: ..................................................................................................................................................700 hPa to 1013 hPa
Expected Service Life: ................................................................................................................................................................................. 3 years
IEC Classification
Medical Equipment
Equipment not suitable for use in presence of flammable anesthetic mixture with air, oxygen or
nitrous oxide.
Type BF, Applied Part
Class I
IP2
The harvester components of the CelluTome™ Epidermal Harvesting System are considered Applied
Parts under IEC 60601-1 Third Edition.