SPECIFICATIONS
Specifications subject to change without notice.
Dimensions: .............................................................................................................. 3.5" W x 6.4" H x 2.16" D (8.9 x 16.3 x 5.49 cm)
Weight (with empty canister attached):...............................................................................................................~.64 lbs (~.29 kg)
Pressure Options: ................................................................................................................................. 75 / 125 mmHg (10 / 16.7 kPa)
Therapy Delivery Modes: ..........................................................................................................................................Continuous or DPC
Canister Volume: ....................................................................................................................................................................................~250 mL
Electrical:
Battery Run Life: ................................................................................................................... ~8.5 hours, depending on settings
Battery Charge Time:..................................................................................................~6 hours from a fully discharged state
External Power Supply Input: ........................................................................................100-240VAC 0.5A - 0.3A 50 - 60 Hz
External Power Supply Output: ............................................................................................................................................... 5V, 2.0A
Patient & Enclosure Leakage Current: ............................................................................................................ <100 Microamps
Environmental Conditions:
Storage / Transport Conditions
Temperature Range: ......................................................................................................................-20.2°F (-29°C) to 140°F (60°C)
Relative Humidity Range: ....................................................................................................................... 0-95% non-condensing
Operating Conditions
Temperature Range: ................................................................................................................................41°F (5°C) to 104°F (40°C)
Relative Humidity Range: ...................................................................................................................15 - 93% non-condensing
Atmospheric Pressure: ............................................1060 hpa (-1253 ft / -381. 9 m) to 700 hpa (9878 ft / 3010 m)
Expected Service Life.....................................................................................................................................................................7.5 days
IEC Classification
Medical Equipment
Type BF, Applied Part
Class II
IP22 - Protection against solid objects greater than 12.5 mm and against liquid water falling for short
periods of time.
The dressing components of the V.A.C.VIA™ Therapy System are considered Applied Parts under IEC
60601-1.
Conforms to AAMI ES60601-1:2005 +A1:2012, IEC 60601-1-6 Ed.3.1:2013, IEC 60601-1-8 Ed.2.1:2012, IEC
60601-1-11 Ed.2.0:2015, CSA C22.2#60601-1:2014 Ed.3.
29