10. Press Start / Pause button to begin the suction microdome raising process. The Status Indicator
light will flash green. If needed, press the Start / Pause button again to momentarily pause the
suction microdome raising process.
11. Once the timer beeps, check the microdomes to see if they are all
ready to harvest - raised, clear, and shiny. (Fig. 1)
12. If the system was paused and microdomes are not ready to
harvest, press the Start / Pause button to continue the process.
13. Once the microdomes are ready to harvest, press the Power
button or the Start / Pause button to stop the microdome
raising process.
14. Unlatch the vacuum head from the harvester. The harvester should remain on the patient's skin.
15. Wipe the underside of the vacuum head following institutional procedures, then place the vacuum
head in the control unit nest and push to engage the latches.
16. Place 3M™ Tegaderm™ Film onto the harvester grid. Center the film starting from the middle then fan
out to the edges to ensure complete coverage of the protruding microdomes.
17. Use fingers to firmly press the 3M™ Tegaderm™ Film against microdomes. Apply firm pressure to
ensure that the 3M™ Tegaderm™ Film adheres to the microdomes.
18. Harvest the microdomes by raising the blue harvester handle fully (until an audible click is heard) and
then lowering the handle to the original position.
19. Carefully peel back the 3M™ Tegaderm™ Film from one end of the harvester. Use care to ensure that
the 3M™ Tegaderm™ Film does not get folded, torn or otherwise damaged when removing it from
the harvester. The 3M™ Tegaderm™ Film with attached micrografts is now ready for application and
should be applied within two minutes of procurement.
20. Apply the 3M™ Tegaderm™ Film with micrografts to the recipient site.
21. Apply additional pressure dressings / bandages as appropriate for the recipient site per your current
standard of care.
22. Remove the harvester from the patient's donor site.
23. Apply a second 3M™ Tegaderm™ Film or other treatment / dressing to the donor site per your current
standard of care.