Operating Instructions for the Acquisition Unit
66 37 461 D 3492
D 3492.201.06.27.02 09.2017
13. Carefully and slowly remove the plug with the holder from the mirror
sleeve by downward movement of the holder. Make sure that the
mirror sleeve points upwards not to risk any damage by liquids
entering the inside of the mirror sleeve.
14. Remove the plug from the holder.
15. Dry the plug with a clean lint-free cloth.
16. Prior to each use of the HLD set test the solution for its effectiveness
per manufacturer´s instructions. If not in use rinse container and
holder and store the holder inside the empty container.
17. Dry the whole surface of the mirror sleeve and also the lower part of
the inner surface using a clean soft, lint-free cloth.
18. Store the mirror sleeve in a manner to protect it from contamination
until the time of use.
19. At time of use remove the white protective cap from the camera tip.
20. Carefully re-attach the mirror sleeve and allow it to lock in place (see
„Refitting the mirror sleeve [ → 51]").
8.1.5.9
High-level disinfection of the mirror sleeve (without HLD Set)
In case the HLD Set is not available
NOTICE
Dry heat sterilization and high-level disinfection must not be combined.
1. Pre-cleaning (see „Pre-cleaning of the mirror sleeve [ → 52]").
2. Remove the mirror sleeve from the CEREC Omnicam (see
„Removing the mirror sleeve [ → 51]").
3. Place the camera in its camera cradle.
4. Use the following disinfectant for the high-level disinfection: CIDEX
OPA.
CAUTION
Observe the disinfectant's manufacturer's safety indications!
8 Maintenance
®
57