Proceeding with scanning procedure
1. Activate the foot control or click on the Omnicam icon with the cursor.
 The scanning procedure begins.
2. Proceed with the scanning procedure as described above.
7.4

Directing the camera

CAUTION
After each use
Reprocess the camera after each patient.
➢ Follow the instructions in chapter „CEREC Omnicam [ → 50]" in
order to avoid cross-contamination between patients.
The CEREC Omnicam acquires images which are used during the
ongoing measurement in spatial relation to each other (image
registration).
During the acquisition and then during the ongoing registration process,
a distinctive sound can be heard.
If the registration cannot be implemented, the acquisition flow is
suspended. You are informed of this by means of a sound. This is
different to the sound emitted during successful acquisition. You can
adjust the volume under configuration and select another type of sound
(melody).
IMPORTANT
Registration error
Should a registration error occur, you must return to another acquired
point.
To start with, practice this procedure on the model and then on intraoral
areas.
➢ Move the CEREC Omnicam to a position where a successful
acquisition was taken. A point that has already been acquired in the
occlusal area is best.
 You will be able to hear the sound for registered acquisitions.
➢ Continue the acquisition.
Divide the acquisition into four consecutive sequences:
1. Occlusal
2. Buccal
3. Lingual
4. Proximal