8 Maintenance
8.1 Care, cleaning, disinfection and sterilization
54
NOTICE
It is recommended that you never switch between disinfecting and
sterilizing solutions. However if you do, make sure you thoroughly rinse
the HLD set before switching.
4. Place the white protective cap on the camera tip and place the
camera in the camera cradle.
5. Use the following disinfectant for the high-level disinfection: CIDEX
OPA.
6. Slid the plug into the metal holder.
7. Press the mirror sleeve onto the plug while holding the metal holder
in order to ensure that no particle contamination or fluids penetrate
the inside of the mirror sleeve.
If attached correctly, the plug seals the mirror sleeve watertight.
CAUTION
Observe the disinfectant's manufacturer's safety indications!