4 Safety
4.2 Safety labels
14
Plug connections of external interfaces
CAUTION
Additional devices connected to external interfaces must be tested
according to the relevant standards, e.g.:
EN 60601-1:1990 + A1:1993 + A2:1995 based on IEC 60601-1, EN
60950-1:2001 based on IEC 60950-1:2001, EN61010-1:2001 based on
IEC 61010-1:2001, UL 60601-1 Part1: first edition 2003, UL 60950 third
edition 2000, UL 3101-1 Part 1 first edition 1993).
They must be installed outside of the patient area (a radius of 1.5m
surrounding the patient).
CAUTION
Low voltages are applied to the sockets for connecting external
interfaces.
➢ Do not touch the pins of the connectors.
NOTICE
The externally connected cables must not be subjected to pulling stress.
Dentsply Sirona
Operating Instructions for the Acquisition Unit
66 37 461 D 3492
D 3492.201.06.27.02 09.2017