8 Maintenance
8.1 Care, cleaning, disinfection and sterilization
58
NOTICE
Only place the mirror sleeve into the container and not the entire
CEREC Omnicam.
Ensure that the mirror sleeve is placed upright in the disinfectant (with
the coated sapphire glass down) so that the solution cannot penetrate
the inside of the mirror sleeve.
5. Choose a container. Find an example on the left.
6. Cautiously fill the container to a filling level of 20mm - 25mm (
- 1in) under the small elongated hole (A) of the mirror sleeve when
immersed. A funnel can be used to assist with filling the liquid. Do not
spill any disinfectant. If spillage occurs please follow disinfectant's
manufacturer's safety indications for cleaning.
7. Insert the mirror sleeve in the correct position (coated sapphire glass
down).
8. Leave the mirror sleeve at least 12 minutes in the disinfectant. While
exceeding the recommended immersion time may not cause damage
to the mirror sleeve, the mirror sleeve should be removed as soon as
possible after the recommended time.
9. Remove the mirror sleeve from the container.
10. Wipe off the mirror sleeve. Use a cotton gauze moistened with tap
water of potable water quality that meets Federal Clean Water
Standards for this purpose.
11. Dry the mirror sleeve using a clean soft, lint-free cloth.
12. Store the mirror sleeve in such a way that it is protected from
contamination until the next use.
13. Carefully re-attach the mirror sleeve and allow it to lock in place (see
„Refitting the mirror sleeve [ → 51]").
Dentsply Sirona
Operating Instructions for the Acquisition Unit
66 37 461 D 3492
D 3492.201.06.27.02 09.2017
13
/
in
16