Precision Novi™ Implantable Pulse Generator
6. If desired, secure the IPG in the pocket by suturing through the holes in the IPG header.
CAUTION: Do not suture through the leads or the splitter.
7. Close and dress the wound(s).
IPG Explant or Replacement
1. Turn off the IPG.
2. Surgically open the IPG pocket and withdraw the device. Please try to preserve the integrity
of all components so that complete device assessment can be performed.
3. Loosen the connector set screws to release and remove the leads, extensions, or splitters.
4. For replacement, connect the new IPG following the instructions for "Connecting the Lead,
Extension, Splitter, or Connector to the IPG" on page 8. Or, to terminate therapy,
surgically remove the implanted lead system.
5. Notify Boston Scientific to document the reason for explant or replacement and to arrange
for return of IPG and components.
IPG Battery Life
For information on battery longevity, see the Information for Prescribers manual.
The clinician programmer and the patient remote control provide notice when the IPG battery
has entered Elective Replacement Mode and when the IPG battery has reached its end of
service (EOS). Refer to the Programming Manual and the Remote Control Directions for Use
for additional information.
Failure to replace the IPG may lead to reduced programming capabilities, limited
communication with the stimulator, and loss of stimulation.
Precision Novi™ Implantable Pulse Generator Directions for Use
90930910-04 REV A
10 of 239