3 INSTALLATION
3.2.5

Wiring of a compact version with relays, with cable glands

Lift the transparent lid after having unfastened the screw. Remove the cover of the device by
unscrewing the 4 screws, pass the cables through the cable glands and connect according
to the pin assignment below.
Always seal the unused cable gland using the supplied obstructor to en-
sure the tightness of the device.
Unscrew the cable gland nut, insert the obstructor and screw the nut back
onto the cable gland.
L+ (12-30 VDC)
4-20 mA output
Strap to be removed if the
4-20 mA output is used
Fit here the cables
using the supplied
cable clips
(see § 3.2.3)
1 2 3 4 PE
Connection relay 2
(see ex. fig. 3.9)
Fig. 3.8
Wiring of the 8205, compact version, with relays, with cable glands
Connecting the 4-20 mA output of transmitter 8205 to a PLC and connection example for
the relays, see fig. 3.9.
1 2 3 4
NC
L+
L-
PE
Without
With
Iout
L+
L-
PE
Supply
12..30Vdc
3A/230VAC
PT SENSOR
Relay cables must be fitted here to the board using
the supplied cable clips (see § 3.2.3)
E-15-
PH TRANSMITTER 8205
Earth (internally connected to the earth plug).
Connect the shielding of the power supply cable
Sourcing / sinking mode switch
CURRENT
SOURCE
SINK
PE
pH sensor connection
PH SENSOR
Temperature sensor connection
Connection relay 1 (see ex. fig. 3.9)