3 INSTALLATION

3.1 Installation Guidelines

The transmitter must be calibrated with buf-
fer solutions (see § 4.2) prior to installation.
Before first electrode calibra-
tion, immerse it for 2 hours at
least in buffer solution pH=7 or
in a solution of KCl 3M (223,6
g/l) or in drinking water.
Pressure-Temperature-Diagram
Mind pressure-temperature dependence
according to the respective fitting materials.
P {bar}
Nominal range
10
9
8
7
PVDF
6
PVC + PP
5
4
3
PVC (PN 6)
2
1
0
-30
-50
-10
+10
+30
+50
Installation Guidelines
Mount the compact pH transmitter (or pH
sensor) in vertical position into a horizontal
pipe.
75°
The transmitter must be protec-
ted from the rain, constant heat
radiation and other environmen-
tal influences such as magnetic
fields or direct exposure to
sunlight.
PVDF (PN 6)
PP (PN 6)
+90
+70
+110
+130
T {°C}
75°
E-8-
PH TRANSMITTER 8205
The electrode must continuously be im-
mersed into the measuring fluid in order to
protect it from drying out.

3.1.1 Installing a compact version

The pH transmitter can be easily installed
in pipes using our specially designed fitting
system. Remove protective cap from the
sensor and keep it for storage.
1. The fitting 4 must be installed into the
pipe according to the installation speci-
fications in section 3.1.
2. Insert plastic nut 3 into fitting, and let
plastic ring 2 snap into guide bush 5.
3. Carefully insert the pH transmitter 1 into
the fitting. If installed properly, the trans-
mitter cannot be rotated.
4. Tighten transmitter housing to fitting with
plastic nut 3.
Plastic nut must only be tighte-
ned by hand!
Fig. 3.1 Compact transmitter installation