3 INSTALLATION
3.4.4 Electrical wiring with Power Supply 115/230 VAC (option)
Remove the cover of the transmitter, the power supply board is in the bottom of the
housing. Pull the cable through PG 13,5 and connect according to fig. 3.8.
The connection of the output signals (current, pulse and relays) is identical to the connection
of the 12/30 VDC transmitter version.

SE35 without relay

230 VAC
115 VAC
L1
N
+
mA
-
1 2 3 4 5
Coil
Hall sensor
Sensor connection
Fig. 3.8 Electrical wiring 115/230VAC power supply
W
arning: If the current output 4...20 mA is not used: set the switch 1 in
!
position A (fig. 3.7) and the strap between 1-3 (fig. 3.6).
+
-
6
Strap (cf §3.4.2)
FLOW TRANSMITTER 8035

SE35 with relays

230 VAC
115 VAC
L1
N
-
mA
+
Switch 1 in position A
Coil
Hall sensor
Sensor connection
7
8
E-12-
1 2 3 4 5
6
Strap (cf §3.4.2)
9
10
+
-