Common Characteristic
Digital Port
Max voltage rating:
Pins 1 & 2 as FLT:
Pins 1-7 as outputs:
Pins 1-7 as inputs:
Pin 8:
Acoustic statement
(European Machinery Directive)
Environmental
Operating environment:
Temperature range:
Relative humidity:
Altitude:
Storage temperature:
Regulatory Compliance
EMC:
Safety:
AC Input
Connections:
Phase and range:
Frequency:
Input VA:
Input current per phase:
Efficiency at full power:
Power Factor:
Keysight RP7970 & RP7980 Series Operating and Service Guide
+16.5 Vdc/- 5 Vdc between pins
Maximum low-level output voltage = 0.5 V @ 4 mA
Maximum low-level sink current = 4 mA
Typical high-level leakage current = 1 mA @ 16.5 Vdc
Maximum low-level output voltage = 0.5 V @ 4 mA;
1 V @ 50 mA; 1.75 V @ 100 mA
Maximum low-level sink current = 100 mA
Typical high-level leakage current = 0.8 mA @ 16.5 Vdc
Maximum low-level input voltage = 0.8 V
Minimum high-level input voltage = 2 V
Typical low-level current = 2 mA @ 0 V (internal 2.2k pull-up)
Typical high-level leakage current = 0.12 mA @ 16.5 Vdc
Pin 8 is common (internally connected to chassis ground)
Acoustic noise emission
LpA 80.92 dB at Operator position: 1
LpA 75.57 dB at Bystander position
LpA 62.04 at Idle fan speed
Normal operation mode per ISO 7779
Indoor use, installation category III, pollution degree 2
0°C to 55°C (Maximum continuous current available is derated at 1% of rating per degree C from
40°C to 55°C)
95% or less (non-condensing)
Up to 2000 meters
-30°C to 70°C
Complies with European EMC Directive for test and measurement products
Complies with Australian standard and carries C-Tick mark
This ISM device complies with Canadian ICES-001
Cet appareil ISM est conforme à la norme NMB-001 du Canada
Complies with European Low Voltage Directive and carries the CE mark.
Conforms to US and Canadian safety regulations.
Declarations of Conformity for this product may be downloaded from the Web. Go to
http://www.keysight.com/go/conformity
L1, L2, L3, N, PE
(Wye-type
ac mains wiring with a neutral connection)
L1, L2, L3, PE (Wye-type ac mains wiring (no neutral) after firmware rev B.06.03.985)
3-phase; 380-480 Vac nominal, line-to-line ±10%
3-phase; 200-208 Vac nominal, line-to-line ±10% (RP798xA after firmware rev B.06.03.1003)
50/60 Hz
23 kVA for 20 kW models, 34 kVA for 30 kW models 2
36 A for 20 kW models, 52 A for 30 kW models 2
90%
0.99 @ nominal input and rated power
1 Quick Reference
All Models
and click on "Declarations of Conformity."
39