[SOURce:]VOLTage:MODE FIXed|STEP|LIST|ARB
[SOURce:]VOLTage:MODE?
Sets the transient mode. This determines what happens to the output voltage when the transient
system is initiated and triggered.
FIXed keeps the output voltage at its immediate value.
STEP steps the output to the triggered level when a trigger occurs.
LIST causes the output to follow the list values when a trigger occurs.
ARB causes the output to follow the arbitrary waveform values when a trigger occurs.
Parameter
FIXed|STEP|LIST|ARB, *RST FIXed
Sets the voltage mode to Step: VOLT:MODE STEP
[SOURce:]VOLTage:PROTection[:LEVel] <value>|MIN|MAX
[SOURce:]VOLTage:PROTection[:LEVel]? [MIN|MAX]
Sets the over-voltage protection level.  Units are in volts. If the output voltage exceeds the OVP level,
the output is disabled and the Questionable Condition status register OV bit is set.
Parameter
0 to 120% of rating, *RST 120% of rating
Sets the over-voltage protection to 24 V: VOLT:PROT 24
An over-voltage condition can be cleared with the OUTput:PROTection:CLEar command after the
l
cause of the condition has been removed.
[SOURce:]VOLTage:PROTection:LOW[:LEVel] <value>|MIN|MAX
[SOURce:]VOLTage:PROTection:LOW[:LEVel] ? [MIN|MAX]
Prevents the output from turning on if the output voltage is below the low voltage protection level.
When the low voltage condition is true, the Questionable Status Register UV bit is set.
Parameter
0 to 102% of rating, *RST 0
Sets the protection level to 2 V: VOLT:PROT:LOW 2
[SOURce:]VOLTage:PROTection:LOW:DELay <value>|MIN|MAX
[SOURce:]VOLTage:PROTection:LOW:DELay? [MIN|MAX]
Sets the low-voltage protection delay in seconds. The low voltage protection function will not be
triggered during the delay time. After the delay time has expired, the low voltage protection function
will be active. This can be used to suppress a false trip when low voltage protection has been enabled
Keysight RP7970 & RP7980 Series Operating and Service Guide
Typical Return
FIX, STEP, LIST, or ARB
Typical Return
<protect level>
Typical Return
<low-voltage protect setting>
5 SCPI Programming Reference
237