Parameter
(none)
Store calibration constants into non-volatile memory: CAL:SAVE
CALibrate:STATe 0|OFF|1|ON [,<password>]
CALibrate:STATe?
Enables or disables calibration mode. Calibration mode must be enabled for the instrument to accept
any calibration commands. The first parameter specifies the state. The second optional parameter is
the password.
Parameter
0|OFF|1|ON, *RST OFF
<password> a numeric value up to 15 digits
Disable calibration: CAL:STAT OFF
Enable calibration: CAL:STAT ON [,value]
A password is required if <password> has been set to a non-zero value.
CALibrate:VOLTage[:LEVel] <value>
Calibrates the local voltage programming and measurement.  The value selects the range to calibrate.
Parameter
The maximum voltage of the output range.
Calibrates the voltage of the 20 V range: CAL:VOLT 20
Keysight RP7970 & RP7980 Series Operating and Service Guide
5 SCPI Programming Reference
Typical Return
(none)
Typical Return
0 or 1
(none)
Typical Return
(none)