Télécharger Imprimer la page

Keysight RP7970 Serie Manuel D'utilisation page 192

Masquer les pouces Voir aussi pour RP7970 Serie:

Publicité

Les langues disponibles

Les langues disponibles

5 SCPI Programming Reference
FUNCtion Command
[SOURce:]FUNCtion CURRent|VOLTage
[SOURce:]FUNCtion?
Sets the output regulation - voltage priority or current priority. In voltage priority mode, the output is
controlled by a constant voltage feedback loop, which maintains the output voltage at its
programmed setting. In current priority mode, the output is controlled by a constant current feedback
loop, which maintains the output current at its positive or negative programmed setting.
Refer to
Priority Mode Tutorial
Parameter
CURRent|VOLTage, *RST VOLTage
Sets the output regulation to current priority: FUNC CURR
192
for more information.
Typical Return
CURR or VOLT
Keysight RP7970 & RP7980 Series Operating and Service Guide