VOLTage Subsystem
Voltage commands program the output voltage of the instrument.
[SOURce:]VOLTage[:LEVel][:IMMediate][:AMPLitude] <value>|MIN|MAX
[SOURce:]VOLTage[:LEVel][:IMMediate][:AMPLitude]? [MIN|MAX]
[SOURce:]VOLTage[:LEVel]:TRIGgered[:AMPLitude] <value>|MIN|MAX
[SOURce:]VOLTage[:LEVel]:TRIGgered[:AMPLitude]? [MIN|MAX]
Sets the immediate or triggered voltage level when the output is operating in voltage priority mode.
The triggered level is the value that is transferred to the output when an output Step is triggered. Units
are in volts. The maximum value depends on the voltage rating of the unit.
Parameter
0.1% to 102% of rating, *RST 0.1% of rating
Sets the output voltage to 20 V: VOLT 20
Sets the triggered voltage to 10 V: VOLT:TRIG 10
[SOURce:]VOLTage:BWIDth:LEVel 0 | 1 | 2, <value>|MIN|MAX
[SOURce:]VOLTage:BWIDth:LEVel? 0 | 1 | 2, [MIN|MAX]
Sets the programming corner frequency associated with each compensation range. The value is in
Hertz. The default frequency is optimized for maximum up-programming speed as well as the fastest
transient response time. It can be reduced to compensate for output overshoots.
0 | 1 | 2 (bandwidth range)
10 to 100,000 for all ranges, *RST: 5,000 (comp0), 450 (comp1), 12 (comp2)
Sets the voltage bandwidth range to 1, frequency 10 kHz: VOLT:BWID:LEV 1, 10,000
[SOURce:]VOLTage:BWIDth:RANGe 0 | 1 | 2
[SOURce:]VOLTage:BWIDth:RANGe?
Sets the voltage compensation range. This lets you optimize output response time with capacitive
loads. These compensation modes only apply when the unit is operating in constant voltage (CV), both
in voltage priority and in current priority (when in voltage limit). Refer to
information about the effect that the compensation settings have on the Programming Speed
Characteristics.
0 (High speed/Small capacitive load) – provides the fastest programming speed and transient
response time.
1 (Medium speed/Medium capacitive load) – provides intermediate programming speed and transient
response time. setting is optimized for stability with a wide range of output capacitors.
Keysight RP7970 & RP7980 Series Operating and Service Guide
Parameter
5 SCPI Programming Reference
Typical Return
<voltage level>
Typical Return
0, 1, or 2
<corner frequency>
Set the Output Bandwidth
for
235