5 SCPI Programming Reference
TRIGger Subsystem
Trigger commands control the transient and acquisition subsystems. Refer to
more information.
TRIGger:ACQuire[:IMMediate]
TRIGger:ELOG[:IMMediate]
TRIGger:TRANsient[:IMMediate]
Generates an immediate trigger. This overrides any selected trigger source.
TRIGger:ACQuire triggers the acquisition system.
TRIGger:ELOG triggers the external datalogger.
TRIGger:TRANsient triggers the transient system.
Parameter
(none)
Generate a measurement trigger: TRIG:ACQ
TRIGger:ACQuire:CURRent[:LEVel] <value>|MIN|MAX
TRIGger:ACQuire:CURRent[:LEVel]? [MIN|MAX]
TRIGger:ACQuire:VOLTage[:LEVel] <value>|MIN|MAX
TRIGger:ACQuire:VOLTage[:LEVel]? [MIN|MAX]
Sets the triggered level of the output.  Applies when the measurement trigger source is set to a level. 
Values are specified in either amperes or volts.  The minimum and maximum values depend on the
ratings of the unit.
Parameter
Voltage: 0 to 102% of rating, *RST 0
Current: -102% to 102% of rating
Set the triggered current level to 3 A: TRIG:ACQ:CURR 3
Set the triggered voltage level to 50 V: TRIG:ACQ:VOLT 50
TRIGger:ACQuire:CURRent:SLOPe POSitive|NEGative
TRIGger:ACQuire:CURRent:SLOPe?
TRIGger:ACQuire:VOLTage:SLOPe POSitive|NEGative
TRIGger:ACQuire:VOLTage:SLOPe?
Sets the slope of the signal.  Applies when the measurement trigger source is set to a level. POSitive
specifies a rising slope of the output signal. NEGative specifies a falling slope of the output signal.
232
Typical Return
(none)
Typical Return
<level value>
Keysight RP7970 & RP7980 Series Operating and Service Guide