Télécharger Imprimer la page

Keysight RP7970 Serie Manuel D'utilisation page 289

Masquer les pouces Voir aussi pour RP7970 Serie:

Publicité

Les langues disponibles

Les langues disponibles

5. Use a flat-bladed screwdriver and carefully pry up on the side of the battery indicated by the
arrow.
6. Install the new battery. Make sure that the positive side (+) is facing up. Place the battery under the
small spring clips indicated by the circle, then push down on the opposite end of the battery indic-
ated by the red arrow. The top of the small spring clips should be visible after the battery is seated
(see red circle).
7. Replace the ribbon cables by fully inserting the cables into the connector; then pushing down on
the locking tab to secure the cables.
8. Replace the Control board and top cover when finished. When reassembling, torque all screws to 9
in-lbs.
9. Reset the date and time (see
Keysight RP7970 & RP7980 Series Operating and Service Guide
Clock
Setup).
7 Service and Maintenance
289

Publicité

loading

Ce manuel est également adapté pour:

Rp7980 serieRp7972aRp7973aRp7982aRp7983aRp7984a