CAUTION: If disinfection is required,
water-based disinfectant solutions must be
used with neutral pH (pH7). Alcohol-based
disinfectant solutions and hydrogen peroxide
are contraindicated, because they can
fade the colour and/or damage the plastic
materials. This is also valid for chemical
products such as acetone and alcohol.
8.3
Cleaning the Tank and Cap
The following procedure must be carried out on the tank and cap of the device.
» PREPARATION
1.
Disconnect the tank from the machine
core unit (see
Chapter 7 on page 26
2.
Unscrew the cap from the tank.
» REQUIRED MATERIALS
• Water;
• Cleaning solution (pH 6-9);
• Clean, soft, low-lint cloth;
• Demineralised water.
CAUTION: DO NOT USE as disinfectants::
• very alkaline products (pH > 9);
• Products containing sodium
hypochlorite;
• Products containing hydrogen peroxide;
• Products containing abrasive substances;
CLEANING AND STERILISATION
CAUTION: Do not sterilise the tank and
);
cap in an autoclave. They could be damaged.
• Very acid products (PH < 4);
• Products containing aldehyde, amine
and/or phenols
• Acetone;
• Methylethylketone;
since they may discolour and/or damage the
plastic materials.
The manufacturer disclaims all liability for any
damage caused by the substances mentioned
above. In case of damage caused by those
substances, the Warranty will be void.
37