• The insert must be held in constant
movement. If the insert is blocked it may
cause overheating of the treated part.
It is recommended to use continuous
motions to minimise contact between
the tip and the part. Do not block them
against the tissues so as not to cause
their overheating. It is advisable to
increase the level of irrigation as the
power level increases.
• Leave the ultrasonic vibrations to work,
do not exert excessive pressure on the
inserts during use. Apply slight force on
the insert for optimum performance.
• When the insert is used in interproximal
spaces, do not block the instrument
6
PUMP/FLUSH FUNCTION
The pump/flush function allows the irrigation circuit rinse cycle to be performed.
CAUTION: Pump/flush function.
The "pump/flush" function must be used after
each treatment, before starting the cleaning
and sterilisation procedures.
Fill the tank with water, preferably
demineralised water. Connect the tank to
the appliance;
24
or leverage on the operating part. The
inserts must be left free to vibrate.
• In endodontic root canal therapy, never
operate the files outside of the root
canal to prevent them from breaking.
To prevent breakages, create a smooth
path with a manual endo file and plan
the most direct possible path of access
to minimise any bending of the insert.
Make delicate movements. Examine
the file often for signs of wear. If the file
breaks inside the canal, avoid contact
between the instrument and the broken
file to stop it from being pushed further
it. Do not apply pressure on the insert in
an axial direction.
CAUTION: If the tubes are not rinsed,
the crystallisation of the salts may seriously
damage the appliance.