5.4
Instructions for Use
After having connected all the accessories as described in
as follows:
To fill the irrigation circuit, use the PUMP
function
by
selecting
on the touch keyboard: all the other
selections will be deactivated and the
writing PUMP/FLUSH will flash;
While the writing PUMP/FLUSH is
flashing, press the pedal once and then
release it, PUMP/FLUSH will stop flashing
and the irrigation circuit will start to fill;
As soon as the peristaltic pump starts
working, the entire scale of values in the
"irrigation" section lights up, and during
the passage of the liquid, the irrigation
value switches from 6 to 0;
The cycle can be stopped as soon as liquid
is seen flowing out from the multipiezo,
handpiece by pressing PUMP/FLUSH or
alternatively by pressing on the pedal.
The PUMP function is deactivated, and
the keyboard becomes active again in the
last setting used;
1
PUMP/FLUSH
2
3
4
Chapter 4.3 on page 11
perio/scaler
perio/scaler
restorative
restorative
perio/scaler
perio/scaler
restorative
restorative
perio/scaler
perio/scaler
restorative
restorative
USE
proceed
21