multipiezo
Position the scaler handpiece connector
above a container or sink to catch the
liquid released during the disinfection
cycle;
CAUTION: Before connecting and
using the scaler handpiece, make sure
that the electrical contacts on both
the scaler handpiece and its cord are
perfectly dry.
Select "pump/flush" on the touch
Chapter 6 on page 24)
keyboard (see
All the other selectable options on the
keyboard will be disabled.
NOTE: The "pump/flush" mode can be
deactivated at any time by pressing the
"pump/flush" key again. The keyboard
will be reactivated and configured in the
last setting used.
Press the pedal once and release it to
start the cycle.
The keyboard will stop flashing and the
cleaning cycle will start;
NOTA: the function can be stopped at
any time by pressing "pump/flush" again
or alternatively by pressing the pedal
briefly. The keyboard will be reactivated
and configured in the last setting used.
As soon as the peristaltic pump starts
working, the entire scale of values in the
"irrigation" section lights up, and during
the passage of the liquid, the irrigation
value switches from 6 to 0. The cycle lasts
20 seconds. Once ended, the keyboard
will be reactivated and configured in the
last setting used;
32
6
7
.
8
9
perio/scaler
perio/scaler
restorative
restorative
perio/scaler
perio/scaler
restorative
restorative
20"