Place the handpiece, with or without
insert, above a container for the liquid
that will be released during the cleaning
cycle;
To enter cleaning mode, select "pump/
flush" on the touch keyboard: all the
other selections will be deactivated;
NOTE: The "pump/flush" mode can be
deactivated at any time by pressing the
"pump/flush" key again. The keyboard
will be reactivated and configured in the
last setting used.
To start the "pump/flush" function while
the writing "pump/flush" is flashing,
press the pedal once and release it:
the keyboard will stop flashing and the
cleaning cycle will start;
NOTE: The function can be stopped at
any time by pressing "pump/flush" again
or alternatively by pressing the pedal
briefly. The keyboard will be reactivated
and configured in the last setting used.
As soon as the peristaltic pump starts
working, the entire scale of values in the
"irrigation" section lights up, and during
the passage of the liquid, the irrigation
value switches from 6 to 0. The cycle lasts
20 seconds. Once ended, the keyboard
will be reactivated and configured in the
last setting used.
After cleaning the irrigation circuit, disassemble the individual parts (see
26
) and proceed to clean and sterilise them (see
PUMP/FLUSH FUNCTION
2
3
4
perio/scaler
perio/scaler
restorative
restorative
5
Chapter 8 on page 29
perio/scaler
perio/scaler
restorative
restorative
20"
Chapter 7 on page
).
25