Fig. 16: Emergency release
1
Hole
9.4
Replacing the mains input fuse
1
Fig. 17: Mains input fuse
1
Fuse holder
2
Snap lock
Model
EBA 200
EBA 200
EBA 200 S
EBA 200 S
AB1800en
1.
Look through the window in the lid to ensure that the rotor is sta-
tionary.
2.
Insert the hex key horizontally into the hole (
until the lid opens.
1
3.
Remove the hex key from the hole (
Risk of electric shock due to maintenance and servicing
work on live device.
−
Personnel:
■
Trained user
The mains fuses are located next to the mains switch.
2
The mains switch is in switch position
1.
Disconnect the mains cable from the device plug.
2.
Press the snap locks (
out.
3.
Replace the defective mains input fuses.
2
Only use fuses with the nominal value specified for the type: see the
table below.
4.
Push in the fuse holder (
5.
Reconnect the device to the mains.
Type
1800
1800-01
1802
1802-01
Rev.: 09 / 01.2024
1 ).
WARNING
Disconnect the device from the mains before carrying
out repairs and maintenance.
[O]
2 ) against the fuse holder ( 1 ) and pull them
1 ) until the snap lock engages.
Fuse
T 1.6 AH/250 V
T 3.15 AH/250 V
T 3.15 AH/250 V
T 6.3 AH/250 V
Troubleshooting
1 ) and turn anticlockwise
Order no.
E891
E997
E997
2266
37 / 39